Toho-electronics TTM-509 Instrukcja Użytkownika Strona 43

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 66
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 42
TOHO TOHO ELECTRONICS INC.
DWG. No.
45-4199-E
PAGE
43/66
Character
PV screen
Auxiliary screen
Designation Description Initial value
10

ch

Set a deviation
display range
Thermocouple input
Setting range: 0 to 999 or 0.0 to 999.9
Setting unit: °C
Resistance bulb input
Setting range: 0 to 999 or 0.0 to 999.9 or 0.00 to
999.99
(when the input type is

,

,
or

)
Setting unit: °C
Current and voltage inputs
Setting range: 0 to 99999
The decimal place can be set to a
desired position.
Setting unit: digit

turns off the display of the deviation.

Potentiometer input
Setting range: 0.0 to 999.9
Setting unit: %
11

□□

Set the buzzer
: Buzzer OFF
: ON when a key is operated
: ON when an event occurs
: ON when a key is operated and when an event
occurs